Mitral valve regurgitation (MR) occurs when a flaw in the mitral valve causes backflow of blood into the left atrium. It is possible to reduce the amount of MR by reshaping the mitral annulus which, in turn, changes the configuration of the valve leaflets. A new percutaneouslydelivered device which induces this reshaping is currently under development at Edwards Lifesciences. The device, which is placed in the coronary sinus (CS), is essentially a spring with anchoring stents at each end. This study examines the mechanical behavior of one of the anchoring stents, from its forming from a small diameter tube to its deployment into the coronary sinus. The anchoring stent is modeled using the ABAQUS Superelastic constitutive model, and the CS is modeled using a hyperelastic constitutive model.