Dec 19 2023

T132-2022 Notification regarding BIOVIA Compose and Capture 2022 HF5

BIOVIA Compose and Capture

Program

BIOVIA Compose and Capture

Operating System

All supported operating systems

Description

We are providing this Technical Note to inform you about the controlled availability (CA) release of BIOVIA Compose and Capture 2022 HF5 which includes the following enhancements and fixed defects:

Enhancements

This release of BIOVIA Compose and Capture includes the following enhancements.

User Story

Description

CAPT-7387

Classic Capture is no longer limited to the first 100 material synonyms when looking for a specific recipe synonym during scan container.

PROP-10800

Compose now provides a parameterized GET that will update changes to phrasing made by utilizing the API.

Fixed Defects

This release of BIOVIA Compose and Capture includes the following fixed defects.

Defect

Description

CAPT-7389

Sample ID and Sample Name are sometimes not showing up in Classic Capture when a recipe was executed.

CAPT-5270

Classic Review displays incorrect data verification signature information for ad hoc observations, if the observations are subsequently modified.  

CAPT-6634

Capture allows registering batch in CISPro without collecting all the required parameters

CAPT-6654

When clearing out values during Data Acquisition recollect, the calculations aren't re-run.

CAPT-6778

Using statistical functions such as sum, minimum, mean, etc with optional parameters doesn't calculate correctly with cleared group rows + any one parameter recollected. A workaround with summary functions could be written, but would be overly complicated.

CAPT-7375

Summary expression selecting values on the same step recalculates to null if any optional dependent is cleared

CAPT-7384

Review shows Material Substitution indicator for non-substituted containers when recipe synonym name contains commas

CAPT-7391

For Capture Hub recipes Only the first 100 Synonyms are retrieved from CISPro when scanning a Container

PROP-10153

When using 3DPassport for authentication, if the session expires, unsaved work may be lost.  This situation should only affect development users at this time.  

PROP-10533

The Compose Recipe Approval Report 2020 fails to execute if BIOVIA Pipeline Pilot Server is deployed on Linux.

CAPT-3944

Step Repeat permanently disables the Done button in Form View for the current step and all subsequent steps.

CAPT-4108

In certain circumstances where a recipe was authored to include a parameter template that is not present on the system where the recipe is executed, an actual value may be improperly evaluated as differing from a planned value.  

CAPT-4622

When Task action requires a signature but PEX action does not, invoking action on the execution will result in PEX and task states getting out of sync

CAPT-6062

Slow performance retrieving initial CISPro container data when scanning barcode from Compose Capture

CAPT-6345

Material is not in Recipe execution Step of PEX Report$ when a calculation is part of the Parameter

CAPT-6541

Cannot receive CISPro container from Capture if using Each units

CAPT-6624

In cases where a required parameter's substr expression calculates a result of an empty string, this meets the need of a result and the step can be completable even though nothing is displayed for the parameter value.  

CAPT-6787

Duplicate unit symbols can cause issues with calculations: Unit mismatch error

CAPT-7042

When using °C in some calculations, the software will actually convert to Kelvin and back, adding 273.15 °C to the result.  This can be worked around by subtracting or dividing as necessary.  

CAPT-7177

Cleared sample data rows in Capture may appear as blank rows in Task Plan results in certain cases.  

Resolution                                                                         

BIOVIA Compose and Capture 2022 HF5 was released on 23 September 2022. To get access to this release, please contact BIOVIA Support.

 

How to contact BIOVIA Support

If you have any questions, please contact BIOVIA Support.