We are providing this Technical Note to inform you about the controlled availability (CA) release of BIOVIA Compose and Capture 2022 SP1 HF2 which includes the following enhancements and fixed defects:
Enhancements
This release of BIOVIA Compose and Capture includes the following enhancements.
User Story | Description |
CAPT-7294 | For Data Acquisition Activity Tasks, Data Acquisition does not require that any data be loaded. This allows protocols to be run against attached measurements and included in the results. |
Fixed Defects
This release of BIOVIA Compose and Capture includes the following fixed defects.
Defect | Description |
CAPT-5270 | Classic Review displays incorrect data verification signature information for ad hoc observations, if the observations are subsequently modified. |
CAPT-6634 | Capture allows registering a batch in CISPro without collecting all the required parameters. |
CAPT-6654 | When clearing out values during Data Acquisition recollect, the calculations are not re-run. |
CAPT-7375 | Summary expression, selecting values on the same step recalculates to null if any optional dependent is cleared. |
CAPT-7384 | Review shows Material Substitution indicator for non-substituted containers when recipe synonym name contains commas. |
CAPT-7391 | For Capture Hub recipes, only the first 100 synonyms are retrieved from CISPro when scanning a Container. |
PROP-10153 | When using 3DPassport for authentication, if the session expires, unsaved work may be lost. This situation should only affect development users at this time. |
PROP-10533 | The Compose Recipe Approval Report 2020 fails to execute if BIOVIA Pipeline Pilot Server is deployed on Linux. |
CAPT-3944 | Step Repeat permanently disables the "Done" button in Form View for the current step and all subsequent steps. |
CAPT-4108 | In certain circumstances where a recipe was authored to include a parameter template that is not present on the system where the recipe is executed, an actual value may be improperly evaluated as differing from a planned value. |
CAPT-4622 | When Task action requires a signature but PEX action does not, invoking action on the execution will result in PEX and task states getting out of sync |
CAPT-6062 | Slow performance retrieving initial CISPro container data when scanning barcode from Compose Capture. |
CAPT-6345 | Material is not in Recipe execution step of PEX Report$ when a calculation is part of the Parameter. |
CAPT-6541 | Cannot receive CISPro container from Capture if using "Each" units. |
CAPT-6624 | In cases where a required parameter's substr expression calculates a result of an empty string, this meets the need of a result and the step can be completable even though nothing is displayed for the parameter value. |
CAPT-6787 | Duplicate unit symbols can cause issues with calculations, resulting in Unit Mismatch error. |
CAPT-7042 | When using °C in some calculations, the software will actually convert to Kelvin and back, adding 273.15 °C to the result. Workaround: Subtract or divide as necessary. |
CAPT-7177 | In certain cases, cleared sample data rows in Capture may appear as blank rows in Task Plan results. |
CAPT-7403 | When scanning a container in Capture to get the mapped extended properties data from CISPro, if one of the materials does not have a value defined for the extended property in the CISPro system, Capture populates the information from one of the other repeats if the value is collected. |
CAPT-7406 | In some cases for a recipe with a repeatable step, the Repeat button is not working in the Horizontal view. WORKAROUND: Click repeat once in the Vertical view, then the step is repeatable in the Horizontal view. |
Resolution